Product characteristics |
---|
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 |
Measuring range | |
Process connection | threaded connection M18 x 1,5 internal thread |
Application |
---|
Special feature | Gold-plated contacts |
Measuring element | 1 x Pt 1000; (to DIN EN 60751, class B) |
Media | liquids and gases |
Pressure rating [bar] | 300 |
Minimum installation depth [mm] | 15 |
Electrical data |
---|
Operating voltage [V] | 20...30 DC |
Current consumption [mA] | < 66 |
Min. insulation resistance [M?] | 100; (500 V DC) |
Protection class | III |
Reverse polarity protection | yes |
Power-on delay time [s] | 1.5 |
Integrated watchdog | yes |
Inputs / outputs |
---|
Number of inputs and outputs | Number of digital outputs: 1; Number of analogue outputs: 1 |
Outputs |
---|
Total number of outputs | 2 |
Output signal | switching signal; analogue signal; (configurable) |
Electrical design | PNP |
Number of digital outputs | 1 |
Output function | normally open / normally closed; (parameterisable) |
Max. voltage drop switching output DC [V] | 2 |
Permanent current rating of switching output DC [mA] | 250 |
Number of analogue outputs | 1 |
Analogue current output [mA] | 4...20 |
Max. load [?] | 500 |
Analogue voltage output [V] | 0...10 |
Min. load resistance [?] | 2000 |
Short-circuit protection | yes |
Type of short-circuit protection | pulsed |
Overload protection | yes |
Measuring/setting range |
---|
Probe length L [mm] | 45 |
Measuring range | |
Set point SP | -39.5...125 °C | -39...257 °F |
|
Reset point rP | -40...124.5 °C | -40...256 °F |
|
Analogue start point | |
Analogue end point | |
In steps of | |
Resolution |
---|
Resolution of switching output [K] | 0.5 |
Resolution of analogue output [K] | 0.125 |
Resolution of display [K] | 0.5 |
Accuracy / deviations |
---|
Switch point accuracy [K] | ± (Pt 1000 + 0,2 K) |
Precision analogue output [K] | ± (Pt 1000 + 0,2 K + 0,4%) |
Display accuracy [K] | ± (Pt 1000 + 0,2 K + 1/2 Digit) |
Temperature drift per 10 K [K] | 0.1 |
Response times |
---|
Dynamic response T05 / T09 [s] | 1 / 3; (to DIN EN 60751) |
Max. measuring/display cycle [ms] | 200 |
Software / programming |
---|
Adjustment of the switch point | programming button |
Parameter setting options | hysteresis / window; normally open / normally closed; min/max memory reset; Display unit; zero-point calibration; analogue output selectable and scalable |
Operating conditions |
---|
Ambient temperature [°C] | -25...70 |
Storage temperature [°C] | -40...100 |
Protection | IP 67 |
Tests / approvals |
---|
EMC | DIN EN 61000-4-2 ESD | 4 kV CD / 8 kV AD | EN 61000-4-3 HF radiated | 10 V/m | DIN EN 61000-4-4 Burst | 2 kV | EN 61000-4-6 HF conducted | 10 V |
|
Shock resistance | DIN IEC 68-2-27 | 50 g (11 ms) |
|
Vibration resistance | DIN EN 60068-2-6 | 20 g (10...2000 Hz) |
|
Mechanical data |
---|
Housing | compact type for adapter |
Materials | stainless steel (1.4301 / 304); PBT; PC; EPDM/X; FKM |
Materials (wetted parts) | stainless steel (1.4404 / 316L) |
Process connection | threaded connection M18 x 1,5 internal thread |
Displays / operating elements |
---|
Display | switching status | LED, red | function display | 7-segment LED display | measured values | 7-segment LED display |
|
Remarks |
---|
Remarks | The values for accuracy apply to flowing water. |
|
Pack quantity | 1 pcs. |
Electrical connection |
---|
Connection | Connector: 1 x M12; Contacts: gold-plated |