What are you looking for?

Greisinger GGA570 Atmospheric Oxygen Sensor

BRAND : Greisinger

atmospheric oxygen sensor, closed sensor type with pressure connection

Universal application, diving gas, longlife, with pressure connection, suitable for GMH 569x

GENERAL:

For devices with external pressure port (GMH 5695/3695) this housing is optimal. Especially for systems with high or low pressure or with existing back pressure by flow.

Details

atmospheric oxygen sensor, closed sensor type with pressure connection

Universal application, diving gas, longlife, with pressure connection, suitable for GMH 569x

GENERAL:

For devices with external pressure port (GMH 5695/3695) this housing is optimal. Especially for systems with high or low pressure or with existing back pressure by flow.

APPLICATION:

It can be screwed airtight (Attention: Observe permissible operating pressure!). The device-pressure port is connected to the sensor pressure port. The device measures and compensates for the actual pressure at the sensor.

SPECIFICATIONS:

Sensor element:
GOEL 370
Oxygen-partial pressure probe, mounted in external sensor housing replaceable (temperature sensor mounted in housing)
Specific features:
Long service life, for protective gases with a high O 2 concentration and oxygen content <35 vol.% O 2
Measuring range:
Partial oxygen pressure:
0 ... 1100 hPa O 2
Oxygen concentration:
0.0 ... 100.0% O 2
Response time: T 90
<10 s
Accuracy (at 25 ° C, 1013 hPa)
<2% O 2 ± 0.2% O 2
<25% O 2 ± 0.5% O 2
> 25% O 2 ± 0.5% O2
Scope of supply:
sensor, flow diverter, T-piece