What are you looking for?

Aoip QUICK-CAL 560 Cost-Effective Compact and Fast Dry Block

BRAND : Aoip

The low temperature Quick-Cal 560 covers the temperature range of -12°C to 140°C (in ambient of 20°C). Quick-Cal 560 tester is idealy suited to the less exacting applications where our larger, better specified baths are unnecessary, which makes it perfect for the medical, sterilization and food industries.

It has interchangeable inserts, so you don’t have to buy another calibrator each time you find a new diameter sensor to calibrate. The low temperature Quick-Cal 560 is particularly exciting since it can work from a small 12 V battery, making it truly portable, alternatively a mains supply is available with 100 / 240 V, 50/60 Hz input.

The dry block comes complete with two inserts, one blank and one drilled 6.8 mm diameter 120 mm deep. Alternatively, we can provide pre-drilled inserts to suit your specific ...

Details

The low temperature Quick-Cal 560 covers the temperature range of -12°C to 140°C (in ambient of 20°C). Quick-Cal 560 tester is idealy suited to the less exacting applications where our larger, better specified baths are unnecessary, which makes it perfect for the medical, sterilization and food industries.

It has interchangeable inserts, so you don’t have to buy another calibrator each time you find a new diameter sensor to calibrate. The low temperature Quick-Cal 560 is particularly exciting since it can work from a small 12 V battery, making it truly portable, alternatively a mains supply is available with 100 / 240 V, 50/60 Hz input.

The dry block comes complete with two inserts, one blank and one drilled 6.8 mm diameter 120 mm deep. Alternatively, we can provide pre-drilled inserts to suit your specific needs. Each calibration well is 13 mm diameter and 120 mm deep.

Key features:

  • Low temperature operation: -12ºC (Tamb 20ºC) to 140°C
  • Fast and compact

General specifications

Temperature range

-12°C (Tamb 20°C) to +140°C

Resolution

0.1°C

Accuracy :
Comparison calibration
Against display temperature with a single probe in the 6.8 mm sleeve

±0.1°C to ±0.4°C
±0.4°C

Stability

±0.05°C

Heating time

0°C to 100°C: 9 min

Additional time for best stability

5 min

Calibration volume

Two pockets of: 13 mm x 120 mm depth

Size

65 x 152 x 175 mm

Weight

1.5 kg

Power supply

Switch mode between 15 VDC or 100 / 240 VAC, 50/60 Hz