Aoip CALYS 75 Field Documenting Multifunction Calibrator

BRAND : Aoip

CALYS 75 is a field documenting multifunction calibrator of high accuracy within CALYS range. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument.

Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…), CALYS 75 makes advanced data exploitation and full data traceability easier, as well as quick access to fu...

Details

CALYS 75 is a field documenting multifunction calibrator of high accuracy within CALYS range. It is the perfect tool for advanced process maintenance and use on test bench in all industries. Suitable for all field and lab measurements, it can simultaneously measure and generate over two isolated channels various signals of temperature, resistance, process, pressure and frequency in one single instrument.

Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…), CALYS 75 makes advanced data exploitation and full data traceability easier, as well as quick access to functions by menus.

The instrument simultaneously measures and simulates:

·         Temperature: Up to 0,014% RDG

·         Resistance: Up to 0,012% RDG, 4 K? range

·         Current: Up to 0,0175% RDG, 50 mA range + 24 V loop supply

·         Voltage: Up to 0,013% RDG, 50 V range

·         Frequency: Up to 0,005% RDG, 20 KHz range (10 KHz in simulation)

·         When used with an external pressure module (ref. ACL433), CALYS 75 can measure and simulate pressure (comparison calibration with a pressure pump).


calys60-calibration

Using this user-friendly instrument, calibration tasks can be quickly carried out over the whole process chain. Take the 900 g documenting process calibrator to the field with you during the whole week with 10 calibration procedures stored in the device. Run the procedure after connecting the probes to the instrument (Easy connect system®) and save the results for onsite easy and quick calibration. Back to the office, you can then upload the data on a computer in order to issue customized calibration certificates with dedicated calibration software DATACAL.

 

IP 54, fully protected by an antichoc rubber holster, CALYS 75 integrates "easyconnect" terminals and a wide backlite display that makes it easy to use in any severe or dark conditions.

Easy connection system

CALYS-EasyconnectConnect your probes by simply pushing on the terminal top and insert wires of up to 3 mm or 10 AWG diameter and compensated thermocouple connectors.

Wires are held tight between two brass plates ensuring thermal stability and a very good cold junction compensation for thermocouples.

This system also enables 4 mm banana plugs and security connectors to be connected on the terminal top.

 

 

CALYS series, 4 models from basic use to advanced performances

Specifications

CALYS 50

CALYS 75

CALYS 100

CALYS 150

Top accuracy

200 ppm

130 ppm

50 ppm

Temperature accuracy

Thermocouples (14)
RTDs (12)

0.013% RDG for Tc K
0.012% RDG

0.01% RDG for Tc K
0.01% RDG

0.005% RDG for Tc K
0.006% RDG

DC current
+ Loop supply 24 V

Range
Accuracy

50 mA
0.0175% RDG

100 mA
0.007% RDG

DC voltage

Range
Accuracy

50 V IN / 20 V OUT
0.013% RDG

50 V
0.013% RDG

50 V
0.010% RDG

50 V
0.005% RDG

Frequency

Range
Accuracy

20 KHz IN / 10 KHz OUT
0.005% RDG

100 KHz
0.01% RDG

Resistance

Range
Accuracy

4000 ?
0.012% RDG

4000 ?
0.010% RDG

50 K?
0.006% RDG

Pressure

Range
Accuracy

Relative pressure: 30 bar / Absolute pressure: 1,000 bar
0.05% RDG

Compliance to standards

21 CFR Part 11

NADCAP Heat treatment AMS 2750

Additional functions

Advanced data exploitation: Scaling, relative measurement, simulation of ramps and steps, synthetizer, square root, statistical functions
Transmitter function

Additional functions

Switch test
Calibration of transmitters

Additional functions

Comparison calibration
HART: Digital calibration and data transfer
Calibration of thermistors

Software

DATACAL calibration software for configuration and data management

Memory

10,000 data stored and recalled on screen as curve or list