CALYS 75 is a field documenting multifunction calibrator of high accuracy
within CALYS range. It is the perfect tool for advanced process maintenance and
use on test bench in all industries. Suitable for all field and lab
measurements, it can simultaneously measure and generate over two isolated
channels various signals of temperature, resistance, process, pressure and
frequency in one single instrument.
Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…), CALYS 75 makes advanced data exploitation and full data traceability easier, as well as quick access to fu...
CALYS 75 is a field documenting multifunction calibrator of high accuracy
within CALYS range. It is the perfect tool for advanced process maintenance and
use on test bench in all industries. Suitable for all field and lab
measurements, it can simultaneously measure and generate over two isolated
channels various signals of temperature, resistance, process, pressure and
frequency in one single instrument.
Providing extended functionalities (temperature
simulation, scaling, steps, synthesizer, statistical functions…), CALYS 75
makes advanced data exploitation and full data traceability easier, as well as
quick access to functions by menus.
The instrument simultaneously measures and simulates:
·
Temperature: Up to 0,014% RDG
·
Resistance: Up to 0,012% RDG, 4 K? range
·
Current: Up to 0,0175% RDG, 50 mA range
+ 24 V loop supply
·
Voltage: Up to 0,013% RDG, 50 V range
·
Frequency: Up to 0,005% RDG, 20 KHz
range (10 KHz in simulation)
·
When used with an external pressure
module (ref. ACL433), CALYS 75 can measure and simulate pressure (comparison
calibration with a pressure pump).
Using this user-friendly instrument, calibration tasks can be quickly carried
out over the whole process chain. Take the 900 g documenting process calibrator
to the field with you during the whole week with 10 calibration procedures
stored in the device. Run the procedure after connecting the probes to
the instrument (Easy connect system®) and save the results for onsite easy and
quick calibration. Back to the office, you can then upload the data on a
computer in order to issue customized calibration certificates with
dedicated calibration software DATACAL.
IP 54, fully protected by an antichoc rubber holster, CALYS 75 integrates
"easyconnect" terminals and a wide backlite display that makes it
easy to use in any severe or dark conditions.
Easy connection system
Connect your probes by
simply pushing on the terminal top and insert wires of up to 3 mm or 10 AWG
diameter and compensated thermocouple connectors.
Wires are held tight between two brass plates ensuring thermal stability
and a very good cold junction compensation for thermocouples.
This system also enables 4 mm banana plugs and security connectors to be
connected on the terminal top.
CALYS series, 4 models from basic use to advanced performances
Specifications |
CALYS 50 |
CALYS 75 |
CALYS 100 |
CALYS 150 |
|
Top accuracy |
200 ppm |
130 ppm |
50 ppm |
||
Temperature accuracy |
Thermocouples (14) |
0.013% RDG for Tc K |
0.01% RDG for Tc K |
0.005% RDG for Tc K |
|
DC current |
Range |
50 mA |
100 mA |
||
DC voltage |
Range |
50 V IN / 20 V OUT |
50 V |
50 V |
50 V |
Frequency |
Range |
20 KHz IN / 10 KHz OUT |
100 KHz |
||
Resistance |
Range |
4000 ? |
4000 ? |
50 K? |
|
Pressure |
Range |
Relative pressure: 30 bar / Absolute pressure: 1,000 bar |
|||
Compliance to standards |
21 CFR Part 11 |
||||
NADCAP Heat treatment AMS 2750 |
|||||
Additional functions |
Advanced data exploitation: Scaling, relative measurement, simulation of
ramps and steps, synthetizer, square root, statistical functions |
||||
Additional functions |
Switch test |
||||
Additional functions |
Comparison calibration |
||||
Software |
DATACAL calibration software for configuration and data management |
||||
Memory |
10,000 data stored and recalled on screen as curve or list |